Lasers · All states

Do I need a laser safety officer for my clinic or med spa?

Short answer

It depends on your lasers and your state, but if you use Class 3B or Class 4 lasers the answer is usually yes. The national consensus standards from the American National Standards Institute (ANSI), Z136.1 and Z136.3 for health care, call for a laser safety officer (LSO) where Class 3B or Class 4 lasers are used, and Z136.3 names salons and spas in its scope. The federal Occupational Safety and Health Administration (OSHA) has no laser standard but expects employers to follow consensus standards. Some states, including Texas, Illinois, Florida and Arizona, require an LSO by rule.

What the national standards say

The ANSI Z136 series is the consensus standard of care for laser safety in the United States. ANSI Z136.1 covers lasers in general and calls for a facility using Class 3B or Class 4 lasers to designate an LSO with the authority to monitor and enforce the control of laser hazards.

ANSI Z136.3-2024 covers health care, and ANSI's summary says its scope includes hospitals, surgery centers, medical, dental and veterinary offices, and "non-medical sites such as salons and spas." The 2024 edition adds that a third-party laser provider is not a replacement for an LSO or a laser safety program, and that the LSO performs a hazard analysis for every Class 3B or Class 4 laser.

What OSHA expects

OSHA has no comprehensive laser standard. Its Technical Manual says OSHA can cite employers under the general duty clause and expects them to follow consensus standards such as ANSI Z136.1. The manual says training "is required for Class IIIB and Class IV laser installations."

OSHA also has no specific standard for laser plume. Its respiratory protection and bloodborne pathogens standards may apply.

States that require an LSO or registration

Only some states register lasers or require an LSO by rule. Examples we confirmed:

  • Texas: register Class 3B and Class 4 lasers within 30 days after beginning operation and name an LSO on each application (25 Texas Administrative Code 289.301).
  • Illinois: register each laser installation before operation, list the LSO, and register systems annually (32 Ill. Adm. Code 315.60).
  • Florida: register Class IIIB and Class IV lasers within 30 days after acquisition with an LSO listed (Florida Administrative Code 64E-4.001).
  • Arizona: each registrant designates an LSO who runs the laser radiation protection program (Arizona Administrative Code R9-7-1434).
  • Massachusetts: all Class 3B and Class 4 lasers must be registered with the Department of Public Health, and the designated LSO must provide a laser safety officer training certificate. Registration is filed through the state's Environmental Health eLicensing Portal.

Who may fire the laser is a separate question

Registration and the LSO rules are about the equipment and the safety program. Who may perform treatments is set by each state's medical, nursing and cosmetology laws and boards. Check both. Rules checked November 11, 2026.

Where our training fits

Our Laser Safety Officer course trains the person who runs the laser safety program: laser classes, hazard evaluation and nominal hazard zones, controls, eyewear selection, signs, non-beam hazards such as plume, procedures, training and incident response, built on the American National Standards Institute (ANSI) Z136.1 and Z136.3 standards. It is not a Board of Laser Safety Certified Laser Safety Officer (CLSO) or Certified Medical Laser Safety Officer (CMLSO) credential, a state LSO qualification, an operator license or accredited continuing education.

Our Laser Safety for Operators and Staff course covers safe operation for the people who run or work near the laser: eyewear, the controlled area, patient and staff protection, plume and fire. It is awareness training, not an operator license or certification.

What to do

  1. List every laser and intense pulsed light device, with its class and wavelength.
  2. If any device is Class 3B or Class 4, name an LSO and give that person the authority to stop unsafe use.
  3. Check whether your state registers lasers, and register on time if it does.
  4. Have the LSO do a hazard analysis, set up eyewear and signs, and write procedures.
  5. Train everyone who works in the laser area, and keep records.

Training that helps

Laser Safety Officer (LSO)

Evaluate each laser from its specification sheet, work out hazard zones and eyewear optical density, set control measures and run the program, built on ANSI Z136.1 and Z136.3. Not Board of Laser Safety certification.

8 hours · Self-paced online · $449 per seat

Laser Safety for Operators and Staff

For surgical teams, medical spa and dermatology operators, dental teams and lab users: matching eyewear to each laser, working inside the hazard zone, preventing airway and drape fires, and what to do after a suspected eye exposure.

2.5 hours · Self-paced online · $149 per seat

Sources

Last reviewed November 11, 2026